Sustainably Produced
High-Quality Gloves For All Users

Latex Examination Gloves

Superior Hand Protection for Effective Contamination Control

Dynawell Latex Examination Gloves (Powder-Free) offer superior yet affordable protection for your hands. Being powder-free, each of our gloves has undergone a proprietary polymer coat and surface treatment to ensure easy donning by users.  Dynawell Latex Examination Gloves (Powder-Free) meet global standards for medical gloves. They also fulfil all regulatory requirements for biocompatibility testing.

Features:

  • Specially formulated for superior barrier protection with low allergenic protein content
  • Polymer coat and surface treated for maximum ease of donning
  • Ambidextrous shape for maximum comfort during extended use
  • Biocompatibility tested to ISO 10993 standard

DYNAWELL DISPOSABLE LATEX EXAMINATION GLOVES (POWDER-FREE)
(ITEM CODE: XXXX)
PRODUCT STANDARD SPECIFICATIONS, PERFORMANCE & CONFORMANCE
TypeDisposable, Powder-Free Latex Examination Gloves
MaterialNatural Rubber Latex (NR)
ColourNatural Chlorinated
Design and FeaturesAmbidextrous, Beaded Cuff, Palm Textured
Donning AidChlorinated/Polymer Coated
Powder ContentMax 2.0 milligrams/glove
Protein ContentMax 50 micrograms/dm2
Available SizesXS, S, M, L, XL, XXL
Storage ConditionsRoom conditions, away from direct sunlight
Shelf Life3 years from manufacture date, with the above storage conditions
Packing Style100 pieces/dispenser, 10 dispensers/carton
Performance StandardsASTM D-3578
EN 455-Parts 1, 2 and 3

PHYSICAL DIMENSIONS – 4 MILS, 240 MILLIMETRES/9 INCHES
SizeXSSMLXLXXL
Average width (mm)78±285±595±5106±5113±5120±5
Average weight (gm)5.0±1.05.5±1.06.0±1.06.5±1.07.0±1.07.5±1.0
Length (mm)Min 240
ThicknessCuff
Palm
Finger
0.08±0.02
0.110±0.02
0.120±0.03

PHYSICAL PROPERTIES – ASTM
Parameters
Before AgingAfter Aging
Tensile Strength (MPa)Elongation at break (%)Modulus at 500% (MPa)Tensile Strength (MPa)Elongation at break (%)Modulus at 500% (MPa)
Dynacare’s Results18-22650-7504.0-5.514-18500-7004.0-5.5
ASTM RequirementsMin 18Min 650%Max 5.5Min 14Min 500%N/A

PHYSICAL PROPERTIES – EN
Parameters
Before AgingAfter Aging
Tensile Strength (N)Elongation at Break (%)Tensile Strength (N)Elongation at Break (%)
Dynacare’s Results9.0-11.0N/A6.0-9.0N/A
EN Requirements (median)Min 6NN/AMin 6NN/A

CHEMICAL ANALYSIS
 Protein ContentPowder Content
Dynacare’s ResultsBelow 50 micrograms/dm2Below 2.0 milligrams/glove
ASTM RequirementsMax 50 micrograms/dm2Max 2.0 milligrams/glove

PERFORMANCE
InspectionRelated DefectsInspection LevelAQL
Watertight TestHoles
G-1
1.5
Visual Inspection
Major Defects2.5
Minor Defects4
Physical PropertiesTensile Strength and ElongationS-24
Physical DimensionsMeasurementS-24